Ekelund O, Ekblom K, Somajo S, Pattison-Granberg J, Olsson K, Petersson A
Infect Dis (Lond) - (-) 1-6 [2021-05-29; online 2021-05-29]
The recently launched high-throughput assays for detecting antibodies against SARS-CoV-2 has contributed to the managing strategies for the COVID-19 pandemic. This study aimed to investigate the performance of three high-throughput assays and one rapid lateral flow test relative to regulatory authorities' recommended criteria. A total of 315 samples, including 150 pre-pandemic samples, 152 samples from SARS-CoV-2 RT-PCR positive individuals and 13 potentially cross-reactive samples were analysed with SARS-CoV-2 IgG (Abbott, Abbott Park, IL), Elecsys Anti-SARS-CoV-2 (Roche, Solna, Sweden), LIAISON SARS-CoV-2 S1/S2 IgG (DiaSorin, Saluggia, Italy) and 2019-nCOV IgG/IgM Rapid Test (Dynamiker Biotechnology Co., Tianjin, China). All assays performed with a high level of specificity ranging from 96.7% to 99.3%. Sensitivity differed more between the assays, Roche exhibiting the highest sensitivity of 98.7%. The corresponding figures for Abbott, DiaSorin and Dynamiker Biotechnology were 80.9%, 89.0% and 72.4%, respectively. The results of the evaluated SARS-CoV-2 assays vary considerably, as well as their ability to fulfil the performance criteria proposed by regulatory authorities. Introduction into clinical use in low-prevalent settings, should, therefore, be made with caution.
PubMed 34053400
DOI 10.1080/23744235.2021.1931434
Crossref 10.1080/23744235.2021.1931434